At 2005-02-24 13:36 -0600, Mike Lockwood wrote:
>This brings up a question - what is the minimum amount of astigmatism
>(cylinder) that the Focault test can dependably detect? What amount
>causes a detectable tilt in the knife edge shadow as it passes the center
>of the mirror? I've never seen either of these quantified. (I'm hoping to
>spark a discussion here. If it has been quantified, I apologize for
>wasting the bandwidth.)
I don't think astigmatism can be "quantified" with Foucault, but, as Tex
sez, p. 85, it can be detected "... if the worker is quite familar with
[the shadows], especially for certain orientations of the defect with
respect to the knife edge." If the axis of the astigmatism is at 45° to
the knife edge, the shadow has opposite bulges on top and bottom. Tex
recommends a bench star test. If you visit my web site in the ATM section,
there's simulated images of both whole-mirror Foucault and the bench star
test on a mirror with just passing primary astigmatism (maybe help workers
to become familar...).
>However, many amateurs can't claim this. What's the worst-case scenario
>here? Someone on the list makes a mirror and it tests to 0.95 strehl with
>the Focault test on multiple axes. Assuming reasonable testing accuracy,
>what is the worst-case amount of astigmatism that can be present in the mirror?
If it's primary astigmatism, it's invisible to multi-axis Foucault, so the
worst-case is horrible.
-- Jim Burrows