Hi Ken,
> I am redesigning my mirror cell and ran my numbers through PLOP.
> My Pyrex mirror has an OD of 12.6" and an average thickness of 1.9". If
> I run a 4 point cell I get an error of 2.01375e-06 mm RMS and a support
> ring at .508337 X radius. This is approximately 1/2 the max error of
> 4.20e-06 mm as per the web page user manual. As a reality check, does
> this sound reasonable or have I missed something?
I just verified your result. It's correct:
2 nm RMS surface = 4 nm RMS wavefront
Michael
*******************************************